Reflux Method Acid Value Tester
Reflux Method Acid Value Tester
MT-AD302
The Reflux Method Acid Value Tester is designed for determining the acid value of petroleum products, transformer oils, turbine oils, and light petroleum products. The instrument is manufactured according to GB/T 264, GB/T 7599, GB/T 258, and GB/T 28552 standards.
Introduction
The acid value is an important parameter used to evaluate the oxidation condition, contamination level, and service life of petroleum products. An increase in acid value may indicate oil degradation, oxidation, or contamination, which can negatively affect lubrication performance and equipment reliability.
This Acid Value Tester adopts the reflux titration method, integrating automatic heating, boiling reflux, titration, endpoint detection, waste discharge, cleaning, and result printing into a fully automated testing process.
The system features an 8-inch color touchscreen, RGB color sensor, laser detection technology, and microcomputer control, ensuring high accuracy, good repeatability, and reliable operation.
It is widely used in power plants, petroleum laboratories, transformer oil testing facilities, inspection agencies, and research institutes.
Typical Applications
The Automatic Acid Value Tester is suitable for determining the acid value of:
- Transformer oils
- Turbine oils
- Lubricating oils
- Hydraulic oils
- Light petroleum products
- Insulating oils Industrial oils
Typical application environments include:
- Power generation plants
- Petroleum refineries
- Transformer maintenance companies
- Oil analysis laboratories
- Inspection and certification laboratories
- Research institutes and universities
Key Features
1. Fully Automatic Acid Value Determination
The instrument automatically completes the entire testing process including heating, reflux, titration, endpoint detection, result calculation, waste discharge, and cleaning.
2. Reflux Titration Method
The analyzer adopts the reflux method principle, ensuring accurate acid value determination in accordance with standard testing procedures.
3. Intelligent Touchscreen Operation
An 8-inch Chinese and English color touchscreen provides convenient operation and menu-guided parameter input.
4. High-Precision Endpoint Detection
The detection system uses advanced laser components and a high-precision RGB color sensor for accurate endpoint determination.
5. Automatic Heating and Reflux
A programmed heating bath automatically performs boiling and reflux operations, reducing manual intervention.
6. Automatic Cleaning and Waste Discharge
After each test, the instrument automatically discharges waste liquid and cleans the reaction bottle.
7. Large Data Storage Capacity
The analyzer can store up to 300 historical test records, allowing easy data retrieval and printing.
8. Durable and Corrosion-Resistant Structure
The housing is made of high-quality steel plate with electrostatic spray coating, providing good corrosion resistance and easy cleaning.
Technical Specifications
|
Name 3947_0d22e5-d8> |
MT-AD302 Reflux Method Acid Value Tester 3947_5d0838-ec> |
|
Test Method 3947_14bc1e-ca> |
Reflux Method 3947_ad7c71-3f> |
|
Measurement Range 3947_75420d-04> |
0.001 – 2.0 mgKOH/g 3947_790ecf-f8> |
|
Resolution 3947_9603f9-41> |
0.001 mgKOH/g 3947_527128-7e> |
|
Sample Injection Method 3947_d21f12-14> |
Peristaltic Pump 3947_c26fa2-1b> |
|
Display 3947_ae4da5-d0> |
8-inch Color Touchscreen 3947_a69503-16> |
|
Rated Power 3947_54d49e-64> |
200 W 3947_7f83ae-12> |
|
Ambient Temperature 3947_e6d28e-05> |
5 – 40°C 3947_fcbfac-c5> |
|
Dimensions 3947_b49f18-81> |
570 × 350 × 340 mm 3947_8fd4b6-c0> |
|
Weight 3947_65c322-89> |
20 kg 3947_9ff798-bf> |
|
Power Supply 3947_a6c15e-91> |
AC 220 V ±10 V, 50 Hz 3947_b87121-4e> |
Standards Compliance
This Reflux Method Acid Value Tester complies with the following standards:
- GB/T 264 – Test Method for Acid Number of Petroleum Products
- GB/T 7599 – Determination of Acid Number of In-Service Transformer Oil and Turbine Oil (BTB Method)
- GB/T 258 – Test Method for Acidity of Light Petroleum Products
- GB/T 28552 – Determination of Acid Number of Transformer Oil and Turbine Oil
- ASTM D974 – Standard Test Method for Acid and Base Number by Color-Indicator Titration
- ASTM D664 – Standard Test Method for Acid Number of Petroleum Products by Potentiometric Titration
These standards are widely used for evaluating the acid value and oxidation condition of petroleum products.
Why Choose Our Our Reflux Method Acid Value Tester
Accurate Acid Value Measurement
Provides reliable determination of acid value for petroleum products and insulating oils.
Fully Automated Testing Procedure
Automatic titration, reflux, endpoint detection, cleaning, and waste discharge reduce operator workload.
High-Precision Color Detection System
Laser components and RGB color sensors ensure precise endpoint identification.
Convenient Data Management
Large data storage and printing functions simplify laboratory record keeping.
Ideal for Transformer Oil and Turbine Oil Analysis
Widely used in power industry laboratories and petroleum testing facilities.